Hospital use High quality portable and warming infusion pump with Battery
Product code:Antis-75
Product Description
Main features
1.Reliable and Safe
1.1, Additional set of bubble alarm design(double bubble alarm) ensure more secure infusion.
1.2, Advanced safety design able to stop infusion and alarm even if the CPU damage the medical staff can use it safely.
2.multipurpose
It can be transfused, transfused blood and transported nutrient solution.
Pumping way
Semi-squeeze finger-type peristaltic pump.
Suitable infusion set
Special or quality PVC/TPE infusion sets with an external diameter of 3.4mm~4.5mm.
User options
Three optional mode:
1. Minimal mode: Users can choose one kind of infusion unit Flow-rate(ml/h), easy to use and avoid error;
2. General mode: Users can choose two kind of infusion unit Flow-rate (ml/h) and Drop-rate (d/min);
3. Proficient mode: All flow-rate modes can be used All functions are open for users to use.
4. Optional Heating Function and WIFI Function
Intelligent conversion of ten infusion modes, the same interface display two infusion modes
10 infusion modes:
1. ml/h;
2. d/min;
3. mg/h;
4. ug/h;
5. mg/min;
6. ug/min;
7. mg/kg/h;
8.mg/kg/min;
9. ug/kg/h;
10. ug/kg/min
Parameter range
1. Flow-rate: 1.0~1200.0 ml/h, 0.1ml/h step;
2. Time: 1s~99h59m59s, 1s step;
3. Drop-rate: 1~333 d/min, 1 d/min step;
4. Dose: 0.1~999.9 mg, 0.1mg step;
5. mg/kg/h: 0.001~999.999 mg/kg/h, 0.001 mg/kg/h step;
6.Body-weight: 1~300 kg, 1kg step;
7. KVO rate: 1.0~5.0 ml/h, 0.1 ml/h step;
8. Bolus rate: 50.0~800.0 ml/h, 0.1 ml/h step;
9. Bolus volume: 0.1~99.9 ml/h, 0.1ml step
Infusion precision
Better than ±5% (using a calibrated or a special IV-set).
Alarm
1.Air in the front-end of tube;
2.The front-end bubble sensor fault;
3.Air in the back-end of tube;
4.The back-end bubble sensor fault;
5.Back-end occlusion, 20~140kPa,8 grades of sensitivity;
6. Back-end pressure sensor fault;
7. Abnormal door-opening;
8.Low battery power, battery fault;
9.Completed infusion;
10.AC power fault;
11.Battery temperature is too high.
Data function
1. Memory of calibration: Records the calibration data of tube automatically for application;
2. Memory of parameter: Records the latest infusion parameter for next operation;
3.Historical records: records latest 1000 infusion events.
(what/when/why flow-rate, etc.)
Class , Internal power supply, CF type, IPX4 equipment.
Power supply
1. Ac adapter:100~240V, 50 /60Hz, Built-in rechargeable Li-battery
2. Power consumption:20VA.
3. Working time of battery: 7h@25ml/h.
Display screen
Color display
Weight and size
94mm (w)*142mm (h)*100 mm (t); about 0.95 kg