| Customization: | Available |
|---|---|
| Classification: | Physiological Functions of Diagnosis and Monitoring Equipment |
| Certification: | CE, ISO13485 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
| Item | Specification |
| Infusion Mode | Volume Mode,Time Mode,Drug Mode,Bolus Mode |
| Rate Unit | ml/hx volume/time,d/min |
| IV Sets | Standard single use IV sets (10,15,20,60d/mL, with diameter 3.4 -4.5mm) |
| Flow Rate | LOml/h-1200ml/h; l.Od/min-400d/min |
| Increment | 1.0-99.9ml/h with O.lml/h; 100-1200ml/h with lml/h |
| Deviation | Flow Rate: W±5% (after calibration) |
| VTBI | 0.1 -9999.9 ml |
| Total Volume | 0-9999.9 ml |
| Time | lmin -99h59min |
| Purge Rate | 600ml/h, displaying purge volume |
| KVO | O.lml/h-5.0ml/h, adjustable |
| Keypad | Numeric Keypad, user friendly |
| BOLUS | Bolus function available,flow rate adjustable |
| Night Mode | Night Mode, No operation for 5mins, LCD brightness lower automatically. |
| Occlusion Pressure Detection |
Occlusion pressure detection , 40~ 130KPa ,0~9 levels adjustable, dynamic pressure value display. |
| Air bubble Detection | Ultrasonic detection,>50|iL,Single bubble volume: 50 ~300 1,0-9 levels adjustable. |
| Volume | 0~9 levels adjustable |
| Motor Anti-reverse Function | Anti-reverse function to prevent upstream |
| IV Set | Open System. Can accept all kinds of IV Set after calibration |
| Self-Testing | Self-Testing when start up and during infusing |
| Audio Visual Alarm Message |
Air bubble, occlusion, door open, infusion finish, standby time expired, battery low, battery empty, battery disengaged, AC Power disengaged, motor error, pump error, pressure sensor error, parameter error, system abnormal |
| Screen | 3.2' brightness LCD |
| Power Supply | AC: 100V -240V, 50Hz/60Hz, W 25VA |
| Battery | Rechargeable Lithium battery: DC11.1V, 3000mAh;operating time: N 10 hours @25ml/h, displaying remaining battery |
| Operation Conditon | Temperature:5°C~40°C,Relative humidity:20% -80% Atmospheric pressure:70~106KPa |
| Transportation & Storage | Ternperature:-20°C~55°C,Relative humidity:10% ~ 95% Atmospheric pressure:50~106KPa |
| Classification | Type CF,Class 1 , IPX3,ISO13485,CE,IEC60601 |
| Dimension | 145mm X 150mm X 200mm ; |
| Weight | 1.7kg |