| Customization: | Available |
|---|---|
| Type: | ECG |
| Certification: | CE, ISO13485 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
| Items | Specification |
| Lead | Standard 12 leads |
| Acquisition Mode | Simultaneous 12 leads acquisition |
| Measurement Range | ±5mVpp |
| Input circuit | Floating; Protection circuit against Defibrillator effect |
| Input Impedance | ≥50MΩ |
| Input circuit current | <0.05µA |
| Record mode | Automatic:3CHx4+1R,3CHx4,3CHx2+2CHx3,3CHx2+2CHx3+1R,6CHX2; Manual: 3CH, 2CH, 3CH+1R, 2CH+1R; Rhythm: Any lead selectable |
| Filter | EMG Filter:25 Hz / 30 Hz / 40Hz/75 Hz / 100HZ/ 150Hz DFT Filter: 0.05 Hz/ 0.15 Hz Filter:50 Hz / 60Hz |
| CMRR | >100dB |
| Patient leak current | < 10µA (220V-240V) |
| Frequency Response | 0.05Hz-150Hz (-3dB) |
| Sensitivity | 2.5mm/mV, 5 mm/mV, 10 mm/mV, 20 mm/mV (error:±5%) |
| Anti-baseline Drift | Automatic |
| Time constant | ≥3.2s |
| Noise level | < 15µV p-p |
| Paper speed | 12.5mm/s, 25mm/s, 50mm/s |
| Recording mode | Thermal printing system |
| Paper specification | Roll 110mmx20M |
| Safety Standard | IEC I/CF |
| Sample Rate | Normal: 1000sps/channel |
| Power Supply | AC: 100-240V, 50/60Hz, 30VA-1OOVA DC: 14.8V/2200mAh, built-in lithium battery |